Impact desorption: DESI uses charged droplets impacting the surface to desorb; AFADESI is its air-assisted improved variant. Laser desorption: LDPI uses laser desorption plus photoionization, matrix-free, 2–3 μm (among the leading in the industry). Electrospray plus photoionization: DPI uses desorption electrospray combined with photoionization secondary ionization.
Gas discharge or plasma: DART uses inert gas passing through dielectric barrier discharge to generate excited species that ionize surface molecules; APCI/APPI mostly use corona discharge or vacuum ultraviolet light to ionize LC effluents and have no spatially resolved mechanism themselves.
Only those with a point-by-point scanning mechanism are used for imaging: DESI/AFADESI, LDPI and DPI are typical imaging ambient sources; AP-MALDI, which brings MALDI to ambient pressure, also counts as an imaging source.
DART itself has no scanning mechanism and requires a moving platform to perform 2D scanning; its resolution and uniformity are weaker than dedicated imaging sources, leaning more toward rapid surface screening than fine imaging.
DESI/LDPI/DPI suit spatially resolved analysis of tissue, herbal medicine and material surfaces; DART suits rapid direct analysis of solid surfaces such as packaging, documents and fingerprints; APCI/APPI suit LC-MS effluent (non-imaging) quantification.
Understanding this distinction avoids conceptual mix-ups: ambient ionization is the broader family, while ambient imaging ion sources are the subset of it that has spatial resolution. Neo-Source LDPI/DPI bring matrix-free ambient ionization into mainstream mass spectrometry imaging.
If spatial distribution is needed, choose DESI/LDPI/DPI/AP-MALDI; if rapid surface screening is needed, choose DART; if LC effluent ionization is needed, choose APCI/APPI. Molecule type also constrains the choice (mid-to-low polarity small molecules lean toward APCI/APPI, metabolites or lipids lean toward DESI/LDPI).
For small molecules, matrix-free, single-cell or subcellular needs, LDPI (2–3 μm (among the leading in the industry), low background above m/z≥70) is currently one of the routes with the highest spatial resolution among ambient ionization sources.
To obtain detailed specifications, compatible models, or a quotation for the MSI LDPI / DPI full series imaging ion sources, visit the Neo-Source official website, or contact the official team for compatibility advice tailored to your mass spectrometer (Agilent / SCIEX / Thermo and other mainstream MS).